What is a point calibration of the pH meter?

Any pH meter must be calibrated by pH standard solution to measure the pH value of the sample. For samples with measurement accuracy below 0.1pH, the instrument can be adjusted by a little calibration method. Generally, the pH is 6.86 or pH7.00. Buffer. Some instruments have an accuracy of only 0.2pH or 0.1pH, so the instrument only has a “positioning” adjustment knob. The specific steps are as follows:

(1) Measure the standard buffer temperature, check the table to determine the pHs at this temperature, and adjust the temperature compensation knob to this temperature.

(2) Rinse the electrode with pure water and dry.

(3) After immersing the electrode in the buffer solution, shake it and place it still. After the reading is stable, adjust the positioning knob to make the instrument display the pH value of the standard solution.

(4) Remove the electrode and rinse.

(5) Measure the sample temperature and adjust the pH meter temperature compensation knob to this temperature value.

(6) Immerse the electrode in the sample solution, shake it and place it still, showing a stable reading.

什么是pH计的一点校准

Educational Building Blocks Toys Wholesale. It's never too early to start children's education and encourage creativity and exploration! Our New Blocks Toy will stimulate spatial thinking, fine motor skills and hand-eye coordination of kids. Bright colors and fun shapes Building Toys Blocks stimulate young minds and build motor skills. This Children New Blocks Toy benefits sensory development to boys and girls of 18 months and above.


New Arrival Building Blocks Toys

New Blocks Toy,Building Toys Blocks,Children New Blocks Toy,Kids New Blocks

HAIPENGDA Plastic Toys Co., Ltd. , https://www.hpdtoys.com